Haemocompatibility testing
Haemobile
- Fresh human blood, used within 30 minutes of collection.
Pulsatile flow with physiological (wall) shear stress.- Low background thrombogenicity.
- Low anticoagulant concentration (typically 0.5 - 1.5 IU/mL heparin).
- No blood–air interface (unlike the Chandler loop model).
Basic research
Preclinical testing
Contact between blood and biomaterials activates the coagulation system and triggers an inflammatory response, reflecting the natural host defence against foreign surfaces. If these responses are insufficiently controlled by natural inhibitors, pathological processes may follow, such as microthrombus formation or thrombosis, bleeding complications, haemodynamic instability, fever, oedema and organ injury. These adverse events are most likely during prolonged and intensive contact with foreign materials, as with vascular implants and extracorporeal circulation.
We specialize in assessing blood compatibility, i.e. the specific interactions between (bio)materials and circulating blood. Using our in-house blood circulation model, the Haemobile, we perform in vitro tests with fresh human whole blood to determine the haemocompatibility of materials, devices and device components.
ISO 10993-4
The haemocompatibility of blood-contacting medical devices must be evaluated before clinical use. The International Standard ISO 10993-4 (Biological evaluation of medical devices — Part 4: Selection of tests for interactions with blood) provides guidance on evaluating these properties.
The standard applies to externally communicating devices, which contact the blood path either indirectly (e.g. blood collection and storage systems) or directly (e.g. catheters and extracorporeal circulation systems), and to implant devices (e.g. stents, heart valves and vascular grafts). Tests are grouped into five categories based on the primary processes involved: thrombosis, coagulation, platelets, haematology and the complement system. Together, these cover all relevant aspects of blood activation. Most importantly, testing should simulate clinical conditions as closely as possible.
Our founder, Dr Wim van Oeveren has served on the ISO committee that drafts and revises ISO 10993-4, so we are well placed to help you design and perform the haemocompatibility testing required for CE marking or FDA submission.
References
[1] van Oeveren W. Obstacles in haemocompatibility testing. Scientifica (Cairo) 2013;2013:392584.
[2] Reviakine I, Jung F, Braune S, et al. Stirred, shaken, or stagnant: What goes on at the blood-biomaterial interface. Blood Rev 2017;31:11–21.
[3] Blok SLJ, Engels GE, van Oeveren W. In vitro hemocompatibility testing: The importance of fresh blood. Biointerphases. 2016;11(2):29802.
[4] Engels GE, Blok SLJ, van Oeveren W. In vitro blood flow model with physiological wall shear stress for hemocompatibility testing-An example of coronary stent testing. Biointerphases 2016;11(3):31004.