Interview: What happens when blood meets a foreign surface?

Sysmex, the supplier of our new scanning electron microscope, interviewed Gerwin Engels and Tessa Lindeboom about what happens when blood meets a foreign surface.

Originally published by Rumi Sarmiento on 29 September 2026 · Translated from the Dutch original on sysmex.nl

However smooth a stent, catheter or artificial heart valve may look on the outside, a far more complex story begins as soon as the material comes into contact with blood. At HaemoScan in Groningen, researchers investigate precisely what happens at that moment. To do so, they use fresh human blood, an in-house flow model and, among other instruments, the NANOS tabletop SEM. General Manager Gerwin Engels and Laboratory Manager Tessa Lindeboom explain how to demonstrate whether a medical material is truly compatible with blood.

In the Laboratory

Gerwin Engels and Tessa Lindeboom in the HaemoScan laboratory.

Why is haemocompatibility so important?

Gerwin: An increasing number of conditions are treated with medical devices that come into direct contact with blood. Examples include catheters, stents and total artificial hearts, but also systems in which blood circulates temporarily outside the body, such as a heart–lung machine.

The body may recognise such a surface as foreign. This can activate, for example, coagulation, platelets, inflammatory responses or the complement system. With a catheter, you do not want a thrombus to form that may later detach, as such a clot can cause serious problems elsewhere in the body.

Persistent low-grade activation can also be harmful. An implant may, for instance, continually activate a small number of platelets, eventually lowering the platelet count in the body. This directly disturbs haemostasis. That is why you want to understand as fully as possible, in advance, how a product behaves in contact with blood.

How do you test this?

Gerwin: A key element is our own in vitro flow model, the Haemobile, in which human blood flows past the test material under controlled conditions.

This requires genuinely fresh blood. The blood is usually in the model within about twenty minutes of collection. In older blood, the function of platelets, among other components, declines significantly, which makes differences between materials harder to demonstrate.

For each donor, we compare different materials using the same blood. In addition, we include a negative control to determine the background response and a positive control known to elicit a clear response. This allows us to minimise the influence of inter-donor variation.

Tessa: Planning sometimes starts very simply, with a message in our donor app group: who can come in next Wednesday? Our donors’ blood is then collected on site, and immediately afterwards we take it to the laboratory.

Microscopic images

Microscopic images of venous stents after exposure to human blood: uncoated (A), with antithrombogenic coating (B), titanium reference (C) and PVC reference (D). The coating markedly reduces thrombus formation.

Why must the blood be tested so quickly?

Gerwin: Blood is not a static test material. Its properties begin to change from the moment of collection. Fresh blood is particularly important for studying platelet activation.

Blood from a blood bank is usually not fresh enough for this type of study. If the platelets no longer respond, or respond poorly, differences between two materials also become less apparent, and you might wrongly conclude that the materials perform comparably.

We also try to prevent activation as far as possible during collection. We therefore use syringes rather than vacuum tubes. The blood is drawn gently and immediately mixed with a small amount of anticoagulant. After all, the response we measure later must be caused by the test material, not by overly aggressive blood collection.

Surfaces of medical devices

The various surfaces of medical devices.

Does every donor’s blood respond in the same way?

Gerwin: No, and that is an interesting point. One donor may respond more strongly to a material than another. That is why we test the materials side by side for each donor. Within a single test, we therefore always compare the same blood, which allows us to control better for inter-donor differences.

At the same time, this raises a broader scientific question. Research often uses blood from young, healthy donors, such as students. In practice, however, an implant is frequently used in an older patient with a medical condition, on medication or with a less healthy lifestyle. Such a patient’s blood may respond differently.

In the longer term, it would therefore be valuable to gain a better understanding of which materials are best suited to different patient profiles. This does make research and product development more complex. Biologically, however, it makes sense that not every body responds to a material in exactly the same way.

What happens after the blood has flowed past the material?

Gerwin: After one hour, we stop the circulation. We then examine what has happened in the blood and on the material.

In the blood, we measure markers of coagulation, platelet activation, inflammation, complement activation and haemolysis, among others. The material is first inspected with the naked eye and under a light microscope; for example, we determine how much thrombus has formed. Fibrin binding and platelet adhesion to the surface are quantified.

We then use the electron microscope, which reveals the surface in much greater detail. For instance, we can distinguish whether platelets are merely adhering to the material or have actually been activated.

The microscopy complements the quantitative measurements. A number tells you how much activation has occurred; the image helps you understand what that looks like on the material.

Why did you need a new electron microscope?

Gerwin: Our old tabletop SEM had been in service for more than ten years. I had replaced parts and carried out several repairs myself, but eventually the image quality became inadequate and the system could no longer be properly supported.

We therefore approached several suppliers. Sysmex responded quickly and came to demonstrate the NANOS at our facility. That was important, because only with your own samples can you really see whether a system is suitable for your application.

What appealed to us was its practical design. There are no moving parts in the vacuum chamber, and the electron source can be replaced by the user. The ability to tilt a sample is also useful, for example when we want to examine the inner surface of a curved material.

Reviewing SEM images

Gerwin Engels reviews the analysed SEM images.

“It is not just about a nicer image. It is about being able to substantiate our results better and present them more clearly.”

— Gerwin Engels

Positioning sample

Tessa Lindeboom carefully positions a sample on the sample stage.

Finally, what difference does it make in your day-to-day work?

Tessa: The system is easy to understand. During the demonstration, we were able to work with it ourselves almost immediately. As a result, once the system had been installed, we could get started quickly.

Gerwin: The software in particular saves us time. We sometimes acquire hundreds of images. With our old system, we had to work out afterwards which image belonged to which sample and time point, and then rename all the files manually.

Now we can add that information directly when saving. Functions that allow a larger area to be captured in sharp focus are also useful for our samples.


About HaemoScan

HaemoScan is an independent contract research organization (CRO) with specialized facilities for haemocompatibility testing and biomarker analysis. It performs preclinical in vitro testing of blood-contacting medical devices and of new drugs in development. In doing so, it helps companies gain deeper insight into their products and supports them in obtaining the regulatory approval needed to bring their products to market more quickly.


Source: https://www.sysmex.nl/academy/knowledge-centre/sysmex-blogs/industry-and-research/wat-gebeurt-er-als-bloed-een-vreemd-oppervlak-ontmoet/